RTDs / Probe RTDs : Series 22
Series 22: Explosion Proof Platinum RTD Probes
22B, 22D
TYPE 22B
RIGID PROBE ASSEMBLIES WITH HEX FITTING
1/4” diameter sheath, available in 316 stainless steel or Inconel
welded to threaded hex fitting for connection head assembly. Operating ranges: –320°F to 900°F for stainless steel and up to 1200°F for Inconel. Teflon® insulated lead wires. single or dual elements; Reference or IEC/DIN R vs. T characteristics and 2, 3, 4 wire element configurations.
TYPE 22D
SPRING - LOADED PROBE ASSEMBLIES
1/4” diameter sheath, available in 316 stainless steel or Onconel; operating ranges: –320°F to 900°F for stainless steel and up to 1200°F for Inconel. RdF’s fail-safe spring-load firs all applications world-wide.single or dual elements.
Reference or IEC/DIN R vs. T characteristics and 2, 3, or 4 wire element configurations.
Thermowell must be specified for this probe.
For more information, visit our Accessories page.
|
Features Calibrations Thermowells & Connection Heads Operating Temperature Range Element Resistance:Ohms Temperature Coefficients |
Accuracy Stability Response Time of Model 22B Vibration Self Heating Recommended Current Insulation Resistance Pressure Rating Lead Material |
||||
|
Teflon® is a registered trademark of the DuPont Corporation. |
|||||
Performance Specifications
RdF explosion proof assemblies are approved for installations in Class I Div.1 Groups B, C, D; Class II Div. 1 Groups E, F, G; and Class III hazardous locations. The spring-loaded or fixed sheathed platinum RTD units suit applications worldwide. Multiple standard options enable the user to customize a configuration for new or replacement applications. RdF’s superior proprietory stainless steel spring-load ensures fail-safe thermal contact while preventing any sheath rotation during vibration.
Three basic styles of construction:
(1) Fixed sheath (2) Spring-loaded sheath (3) Spring-loaded sheath with lagging.
Detailed dimensions for thermowells are here.
Ordering Specifications
|
22B 22D |
Rigid Explosion Proof Platinum RTD Assembly with Connection Head Spring-Loaded Explosion Proof Platinum RTD Assembly with Connection Head |
||||||||||||||||||||||||||||||
| RTD Resistance Specification (ohms @ O°C) | |||||||||||||||||||||||||||||||
|
10 11 T01 20 21 T05 T10 102 112 T012 |
100 100 100 200 200 500 1000 100 100 100 |
± 0.1 ± 0.1 ± 0.12 ± 0.2 ± 0.2 ± 0.6 ± 1.2 ± 0.1 ± 0.1 ± 0.12 |
(α = 0.003924 Ω/Ω/°C) Reference Standard (α = 0.00385 Ω/Ω/°C) IEC/DIN Standard (α = 0.00385 Ω/Ω/°C) Thin Film (α = 0.003924 Ω/Ω/°C) (α = 0.00385 Ω/Ω/°C) (α = 0.00385 Ω/Ω/°C) Thin Film (α = 0.00385 Ω/Ω/°C) Thin Film (α = 0.003924 Ω/Ω/°C) Dual Element (α = 0.00385 Ω/Ω/°C) Dual Element (α= 0.00385 Ω/Ω/°C) Thin Film, Dual Element |
||||||||||||||||||||||||||||
| 4 Sheath Diameter 1/4" only | |||||||||||||||||||||||||||||||
| S Sheath Material – S = 316 SS (Thin Film in S only), I = Inconel | |||||||||||||||||||||||||||||||
| Service Parameter – Sinusoidal Vibration 20Hz to 2kHz | |||||||||||||||||||||||||||||||
|
S H |
Standard construction (25g wire-wound, 80g thin film) |
||||||||||||||||||||||||||||||
|
Probe Length 22B – “L” in inches - see drawing (3" min., 4" min. on heavy duty construction) Stem Length 22D – “A” in inches - see drawing (2" min., 3" min. on heavy duty, X + A = 24" max.) |
|||||||||||||||||||||||||||||||
| Number of Leads | |||||||||||||||||||||||||||||||
|
A B C D E |
Single element, 2 wires Single element, 3 wires Single element, 4 wires Dual element, 2 wires per element Dual element, 3 wires per element |
||||||||||||||||||||||||||||||
| Lead Length in Inches (4" STD.) | |||||||||||||||||||||||||||||||
| Connection Head | |||||||||||||||||||||||||||||||
|
G |
Explosion proof cast aluminum (up to 250°F ambient) |
||||||||||||||||||||||||||||||
|
H |
Explosion proof cast aluminum (up to 500°F ambient) |
||||||||||||||||||||||||||||||
| X |
Explosion proof cast aluminum (up to 250°F) order separately with transmitter (4 to 20 mA) |
||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||
|
– |
|
– |
|
– |
|
– |
|
– |
|
– |
|
– |
|
– |
|
— | – | Typical write-up for 22B | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
– |
|
– |
|
– |
|
– |
|
– |
|
– |
|
– |
|
– |
|
– |
|